Coagulation analyzer BC1, i-CHROMA, HemoChroma, Hemostasis Reagent, INR Simple Simon

 Deutsch

 You are here Products   One-Channel Ball Coagulometer BC1

 

SYCOmed

  

 

Information

 
 
 
 
+49 (0)5261.281 96 62
+49 (0)5261.943 27 28
info@sycomed.com

 

BC1 Ball Coagulometer

Flexible | Versatile | Reliable

The patient’s sample and a test related reagent are included in a diagonally bedded cuvette that slowly turns around its center line. Due to its gravitational force, a metal ball in the reagent mixture moves exactly to a specified spot prior to coagulation. Upon coagulation, the ball is taken along by the fibrin fibers that build-up. This change of position generates the signal in an electro-magnetic sensor which stops a timer set beforehand.

Designed for small sample appearances in a laboratory, this handy and user-friendly analysis device is also used in research and veterinary diagnostics.

The
BC1 coagulometer is reagent independent. All common clot reagents available in the market can be used.

Using suitable test reagents, both plasma and whole blood samples can be measured.

BC1 one-channel coagulometer

BC1 one-channel coagulometer

Illuminated two-line graphic display

Illuminated two-line graphic display


As exclusively the fibrin formation is captured by an electro-magnetic sensor, the clouding of the particular sample has no impact.

The beginning of the measurement is either automatically launched by adding the reagent with an automatic pipette or manual activation of the start button using any common laboratory pipette.

The measurement results evaluated are shown on a high resolution, clearly visible display in seconds, % PT, INR or g/L for fibrinogen determination.
Optionally a low-noise thermo printer can be provided for documentation of the measuring results.

As an option, the
BC1-MACRO version or the BC1-MICRO version for bisection of the test volume are available.

All device versions are optionally available with an automatic 3 volume pipette with 50, 100, and 200 µl pipetting volume.

BC1 Set consisting of thermo printer, automatic 3 volume 
                  pipette, ball dispenser and coagulometer

BC1 Set consisting of thermo printer, automatic 3 volume pipette, ball dispenser and coagulometer

 

Specifications BC1

   
Measuring channels: 1
Test method: electro-mechanical, by ball
Model: optional in MICRO or MACRO Version
   
Sample: hemolytic, icteric and lipemic plasmas, whole blood,
citrated whole blood
Sample volume: >20 µl ( 10 µl )
Reagent: open System for all Clotting Tests ( fibrin formation )
Test volume: 75 - 600 µl
Result output: Seconds, % PT, INR, g/L Fibrinogen
Incubation stations: 2 also useable for sample preparation, tempered at 37°C
Incubationstimer: 0 up to 999 Seconds
Reagent stations: 2 ( 11,8 and 15 mm diameter ), tempered at 37°C
Calibration curve storage: 2 freely programmable calibration curves with 3 points with
% and g/L evaluation
Interface:: RS 232
Thermo printer or PC: via RS 232
Service: maintenance free
Languages: German, English, French, Russian,
additional languages on request
Display: 2-line high resolution graphic display
   
Power supply voltage: Wide range power supply 100 - 240 volts, 50-60 Hz
Operating voltage: 12 Volts- ( also useable in a car )
Electric power: max. 20 VA
Width: 125 mm
Height: 72 mm
Depth: 216 mm
Weight: 1,2 kg ( 2,5 lb )
   
 

Coagulometer

 

 
 
 
 
 
 
 

BC1 Leaflet

 

 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Copyright © 2009-2010 SYCOmed e.K. All Rights Reserved.